[Page 1] Quick Manual Produced by SAMSUNG MEDISON Global Clinical Group ※ Quick Manual dose NOT include all instruction. For more detail, please refer to HS30 User Manual. SAMSUNG MEDISON CO., LTD.
[Page 2] Power Setup Turns system on/off The general system setting Patient Mode Enters, Searches, or Starts or finishes each Changes patient Mode (2D, C, PD, PW, M) information Adjusts the gain control SonoView Q Scan SonoView is a image Auto image optimization management program Toggle control of Monitor menu Zoom Report Changes the parameters on each screen Magnifies an image Shows the measurement (Rotation and Push the button) results of the current application and other Depth / Focus information Adjusts scanning depth and focal point End exam Finishes the exam of the Change key currently selected patient and resets the related data Changes the trackball functions (i.e. Color box, SV, size, etc.) Select Probe Calculator & Caliper TGC Starts measurements by Time Gain Compensation application Measures Distance, Area, Volume, Velocity Menu Opens the chosen left menu on the monitor Dual Left, Right key separation Pointer Selects soft menu or Freeze uses for image review of Pauses/Resumes scanning thumbnail U1 ~ U6 Clear Set (Exit) User customized keys Selects an item or value using the trackball Deletes Text, Indicator, (Exits the function currently being used and (i.e. Save, Print, Clip Store, BodyMarker returns to the previous state) Annotation, Full screen, etc.) I 2
[Page 3] Worklist Search Registration of Patient ■ Select [Patient] on the control panel. ■ Select [Patient] on the control panel then select [Patient]( ) in ( Worklist search is available only when DICOM is connected.) the monitor screen. Search When you click [Search] the list of patients who match the search Insert ID Enter patient ID, name, and other types of basic patient. criteria will be displayed. (Patient list update) Study Add required information for individual applications. ID Select Select the patient list and double click. Information * Enter the OB information. This applies the selected patient information to the system. Application OB tab → LMP, EDD, Fetuses No, etc. Start Exam Click [OK] applied the patient information to the system and Operator Enter name of the physician who scanned the patient. switches the system into scanning mode. Start Exam Click [OK] to enter scan mode. Searching Search stored exam in the system. Patient The selected patient information will be applied Information to the system. I 3
[Page 4] Information Area Probe Selection Quick Preset Changes both probes and presets simultaneously. Probe Select the [Probe] button on the control panel. Shows the frequently used presets. Selection Select [Quick Preset] by using [Pointer] button or [Menu] knob button. Probe Type Displays a list all probes currently connected to the system. (Select up to 8 presets on setup menu) Selects the probe icon you would like to use. * Setup → System → Quick Preset User Preset User presets function can be set up on user’s preference. L/R Flip Left/Right Flip U/D Flip Up/Down Flip Default Preset Displays a list of default applications the selected probe supports. Default presets are factory setting that was Layout Single, Quad image optimized on each probes and applications. Keymap Setting for user defined keys, including the positions of [Set] Preset list Displays a list of presets that the selected application supports. and [Exit] button. When you click , you can use User Preset list. ⁎ Setup → Customize → Buttons When you click , you can use Default Preset list. I 4
[Page 5] Creating a User Preset Modifying a User Preset ■ Function that can create User preset ■ Function that can modify current preset Probe Select Select the [Probe] button on the control panel. Adjusting Adjust image parameters that user thinks proper for Parameters overwriting it. Preset Select Select Default preset what you want to change image [Default] parameters. Probe Select Select the [Probe] button on the control panel. Adjusting After adjusting image parameters that user thinks proper, Edit Press the [Edit] button on the left side of the monitor. Parameters for saving it, select [Probe]button on the control panel. Overwrite Select [Overwrite] button. Preset Select This enables you to only overwrite the currently selected Select [User Presets]. [User Presets] user preset. [+] Select Selecting the [+] button opens the Create New User Preset Rename Changes the user preset’s name. window. Enter the name you want to use and press [OK] button. Delete Deletes the user preset that you select. I 5
[Page 6] 2D Mode Harmonic Provides the Optimal Harmonic Imaging high frequency function to optimize images. S-Harmonic Provides clearer image from near to far zone using wide band frequency. Dual Live Displays 2D image and Color Doppler image simultaneously. Trapezoid Changes rectangular format from linear probe to trapezoidal To use Top menu - Rotate the Soft Menu dial-button format, so can provide wider view angle. (Only for Linear probe) ElastoScan Provides elasticity imaging of an object. (Optional feature) To use Sub menu - Press the Soft Menu dial-button Panoramic Acquires wider image range via reconstructing multiple frames. (Optional feature) Needle Mate+ Provides better needle visibility. (Optional feature) Enter [Full Screen] mode. assign [Full Screen] mode to one of the User Keys. Frequency Enables you to set the probe’s frequency. The selected frequency is displayed the images area on the monitor. (Res: Resolution, Gen: General, Pen: Penetration ) DR Adjust the contrast value by changing the ratio of the minimum and maximum values of input signals. The higher value provides smoother image quality. Reject Reduces noise or echoes from an image in order to make the image clearer. ClearVision Provides clearer tissue boundaries using the noise reduction filter and generates sharp 2D images. (Select up to 5 levels) Frame Average Use this to minimize the appearance of speckles in updated images when the same location is scanned repeatedly. Gray Map Change the 2D Post Curve. The higher value provides stronger contrast image quality. (Select up to 12 levels) MultiVision Combines complex beam steering and scan lines to provide the improved resolution. (Select up to 3 levels) Line Density Set the scan line density. Selecting High increases the number of scan lines and improves the image resolution. (Next Page) Biopsy Guide Adjusts the biopsy guideline before using the biopsy feature. (Next Page) Edge Enhance Allows you to view more accurate images of organ or tissue boundaries. The higher value provides more accurate images of boundaries. (Next Page) I 6
[Page 7] Color/Power Doppler/S-Flow PW/CW Doppler Scale Adjusts the PRF. (Pulse Repetition Frequency) Scale Adjusts the PW/CW PRF. (Pulse Repetition Frequency) Street Invert Inverts the direction of color box. Baseline Adjusts baseline by rotating dial-button. Steer Angle Adjusts the angle of color box. (-8, -4, 0, 4, 8) SV Size Selects a SV size between 0.5~20mm. Color Automatically adjust the appropriate color scale. Auto Calc. Automatically calculate the values of Doppler. (Live On) S-Flow High sensitivity bi-directional Power Doppler mode for tiny vessels. (Access from PD mode → S-Flow mode) Simultaneous Views real-time 2D and Spectral Doppler image at the same time. Auto Gain Smart Auto Doppler * This function is only available in Vascular preset of linear probe. Select [Auto Gain] on the touch screen in Artery, Carotid preset. Select [Quick Scan] on the control panel in the Doppler mode. - Adjusts proper Color gain automatically frame by frame. Location of the ROI box, the sample volume position, and the direction of the steer - The dynamic color coding up on blood pressure can be controlled via angle can be controlled by one click. [Auto Gain] function. -Improves the complex UI Workflow of Doppler mode. I 7
[Page 8] Measurement ■ Caliper Measures (using the trackball and [Set] button) selected items. (i.e. Distance, Area, Volume, Doppler measure etc.) Move the cursor with the trackball or the [Menu] knob button , and press the [Set] button or the [Menu] knob button. Distance Distance Measures the straight distance between 2 points. Velocity Velocity Measures the velocity at 1 specified point. Trace Length Measures traced curved-line. Accel* Measures the velocity to calculate the time and %Stenosis(D) Diameter of a vessel is measured and stenosis ratio acceleration. calculated. Time(D) Calculates the time between the bars. Area Ellipse Measures the circumference area. HR Measures the Heart rate between the two bars. Trace Measures the trace and area of an irregular object. %Stenosis(A) Area of a vessel is measured and the stenosis ratio is Ratio RI Measures the RI (Resistivity Index) at 2 specified points. calculated. Trace Auto Automatic spectrum trace within whole range. Default Shows the frequently used calipers. Limited Automatic spectrum trace within specified range. Tool Use Default Tool by pressing [Change Key] button on the control panel. Manual Trace Spectral Doppler wave form and calculate velocity, and Select up to 3 calipers on each image mode. (2D Mode, D Mode, M Mode) the average velocity of blood flow. (includes S/D ratio, RI, PI) ⁎ Setup → Measurement → Application Accel* : Acceleration I 8
[Page 9] Measurement ■ Calculator Data measurement per each application. (OB, Cardiac, Vascular, Prostate, etc.) Move the cursor with the trackball or the [Menu] knob button , and press the [Set] button or the [Menu] knob button. Application Provides calculate package up on selected probes and Trace Auto Automatic spectrum trace within whole range. applications. Limited Automatic spectrum trace within specified range. Manual Trace Spectral Doppler wave form and calculate velocity, and Group Select a measurement group by pressing the [Menu] knob button the average velocity of blood flow. (includes S/D ratio, RI, PI) or using the [Pointer] button. (Customize order of Group on Setup) Velocity PS, ED, MD Measures the velocity at specified 1 point. ⁎ Setup → Measurement → Menu → Group (Peak Systolic Velocity, End Diastolic Velocity, Min Diastolic Velocity) Label Select a measurement label by pressing the [Menu] knob button PI, RI, S/D Measures the velocity at specified 2 points. or using the [Pointer] button. Provides Auto Sequence that executes selected items in a group Volume Vol. Measures the volume flow value by selecting two bars on a in sequential order. Flow Flow(D) Spectral Doppler image, specify two points in 2D Image Mode. (Customize order of Label on Setup) Measures the volume flow value by selecting two bars on a ⁎ Setup → Measurement → Menu → Label Vol. Flow(A) Spectral Doppler image, specify an ellipse in 2D Image Mode. I 9
[Page 10] EzCompare™ ■ [EzCompare™] displays the previous exam image and the current exam image side by side on the split screen, allowing you to compare and assess it. Automatically matches the image settings, annotations, and Body Marker used in the previous study. ID Select Select the ID from the patient list using the trackball and [Set] Clinical Benefit button. Select Using [EzCompare™] makes it possible to compare accurate comparison with Switching to [EzCompare] mode and move to scan mode. [EzCompare] previous tests and checking the difference in variation of tissue. Check Thumbnail ① Fatty Liver diagnostic: Fatty Liver levels can be easily recognizable according to The stored images are displayed as thumbnail and date tab appears. Field the brightness of the Liver. ② MSK examination: MSK testing is difficult to create the same image as the Compare On one side of the screen, selected images are displayed. On the images previously examined because of the deviation of the images due to the Prior study other side, live scan image will be displayed as identical setting scanning angle. of selected image. (Dual screen) ③ A Follow-up : When the follow-up is examined that is useful to observe changed ⁎ Live scan is available only when the same probe/preset as the aspect lesion such as size, shape, or echogenicity, etc. one applied to the saved image is selected. I 10
[Page 11] SonoView – Image Management, Export & Backup Tab [SonoView ] of the control panel. Export Converts image to PC compatible format such as BMP, JPEG, Select [Exam List] of the monitor screen. TIFF, AVI, MPEG, DICOM. Select Drive. ( CD-G:, USB-F: ) Directories – Select the location in which the exam will be saved. Enter the file name. Select file and video format. (BMP, JPEG, TIFF, MPEG, AVI, DICOM) Export Option Hide Patient Information – To remove patient ID and name Create Patient Folder – Categorized files per date and patient ID Click [Export]. Select Exam Check the combo box ( ) in front of the patient ID in the [Exam List]. Send Transfers the exam to PACS server. (Select Send → Displayed the DICOM Storage → Select Alias → Transfer) ⁎ Transfer all images on selected exams. Delete Deletes the exam. Review or Reviews the saved images. Continue Exam Exams performed longer than 24 hours, only review is allowed. Backup Back up and restore data. These backup data only can be review If the exam was performed within 24 hours from now, append on the system. exam is allowed. Select Backup → Confirm [Yes] → Select Drive → DVD(G:), USB(F:) I 11